Thiết bị giám sát độ rung A3800
Giá:Liên Hệ
- Optional number of input channels
- 4 – 16 channels AC
- 4 – 16 channels DC
- 1 – 4 TACHO inputs
- Adaptive algorithm of data acquisition
- Compact size, DIN rail mounting
- Input channels AC:
- 4 – 16 AC, ICP® power supply on/off
input impedance 100kΩ
integration single, double
high pass filter 1 Hz – 12800 Hz
low pass filter 25 Hz – 25600 Hz
- Input channels DC:
- 4 – 16 DC for process values
input impedance 100 kΩ (VDC), 250 Ω (mADC)
- TACHO inputs:
- 1 – 4 independent TACHO for external trigger
speed range 0,8 Hz – 1000 Hz
- Input range:
- AC +/- 12 V peak-peak
DC +/- 24 V or 4 – 20 mA
TACHO +10V
- AD conversion:
- 24 bit, 64 bit double floating point internal signal processing
No AutoGain function!
- Dynamic range S/N:
- 120 dB
- Frequency ranges:
- max. 25,6 kHz (16 Ch, 65,5 kHz sampling)
- Sampling mode:
- Fully simultaneous for group of 4 channels
- FFT resolution:
- Min. 100 lines
Max. 25 600 lines
- Processor:
- Intel Atom E3815 – 1.46 GHz
- Memory RAM:
- 2 GB
- Internal hard drive:
- SSD 32 GB
- Data acquisition:
- Overall values
Time signals
FFT real time analysis
DEMOD – ENVELOPE analysis
ACMT – low speed bearing analysis
order analysis
user band pass analysis
RPM measurement
DC measurement
Orbit measurement
Speed measurement
- Signal Recorder:
- 64 kHz sampling frequency
4 Ch memory consumption 3 GB/hour
4 Ch total recording – 10 hours
- Trigger:
- free run, TACHO, external (voltage)
- Communication:
- Ethernet 1GB RJ45
- Temperature range:
- -10˚C to +50˚C
- Power:
- DC 12 V
- Case:
- aluminium box
- Size & Weight:
- 117 x 117 x 58 mm (without connectors)
800 g