I. LS117 Optical Density Meter Introduction
Linshang LS117 densitometer is an instrument that can measure both optical density and light transmittance. The instrument adopt the diffuse transmission principle which is suitable for the optical density test of aluminized film, film, X-ray film, lens ink and other materials. The light transmittance test of opalescent translucent materials and frosted matte materials, such as zirconia, opalescent acrylic and so on. The light source of the instrument complies with the CIE photopic luminosity function and can pass the inspection of Chinese national metrological institute.
Related knowledge of optical density: optical density, abbreviated as OD. It is defined as the logarithm of the ratio of incident light to transmitted light, or the logarithm of the light transmittance reciprocal. The calculation formula is OD = log10 (incident light / transmitted light) or OD = log10 (1 / light transmittance). Generally, when testing materials with very low light transmittance, it is difficult to compare the differences between the materials using the light transmittance value. The data will be more intuitive after converting into optical density values.
II. LS117 Optical Density Meter Characteristics
1. The optical density meter has wide measuring range. The optical density value can be measured to OD 6. The transmittance resolution is up to 0.005%.
2. During the test, the interface displays both the transmittance value and the optical density value of the material.
two test data of LS117 optical density meter
3. The LS117 portable densitometer uses a diffuse transmission light path design, which has the same effect of the integrating sphere. There are relevant instructions in “GB / T 11500-2008 / ISO 5-2: 2001”. The regulation has stated that “Use opale glass method instead of integrating sphere method as the basis of ISO standard diffuse transmission density”.
4.The light source complies with the CIE photopic luminosity function to ensure that it passes the testing of the Chinese Metrology Institute.
5. The optical density meter has two measurement methods which is more flexible to use.
two measurement methods of LS117 light transmittance meter
(1) Fixed measurement: It is more suitable for testing regular or smaller objects.
(2) Hand-held measurement: take the probe down for testing, which is more suitable for irregular and large-volume materials.
6. Suitable for testing optical density of aluminized film, film, X-ray film, lens ink and other materials. The optical density meter is suitable for testing the light transmittance of opalescent translucent materials, matte materials.
LS117 test optical density value
LS117 light transmittance meter test light transmittance
Thông số kỹ thuật
Light transmittance
measurement accuracy |
±1% (0%-50%) ±2% (50%-100%) |
Light transmittance
resolution |
0.0005% (0%-10%)
0.005% (10% -100%) |
Optical density
measurement range |
0.000OD-6.000OD |
Optical density
measurement accuracy |
±0.02 (0-2.00OD)
±2% (2.00-6.00OD) |
Optical density resolution | 0.001 OD (0.00-2.00OD)
0.01 OD (2.00-4.00OD) 0.1 OD (4.00-6.00OD) |
Aperture | 2mm |
Light Source | 380nm-760nm, conforming to
CIE photopic luminosity function |
Weight | About 576g (without battery) |
Power supply | 4*AAA alkaline dry battery |
Display | 240*160 dot-matrixes LCD |
Dimension | 148mm×76mm×26mm (L*W*H) |